Skip to content
   

Slurry Carts

Built for Pressure. Designed for Precision.
The HPHT Slurry Cart is engineered to mix and deliver drilling fluids across the face of a core sample in the 6100 Formation Response Tester (FRT) core holder at a designated pressure and temperature. This process enables accurate simulation of fluid loss, which is essential in preventing issues such as formation damage, wellbore instability, and lost circulation during drilling operations. With advanced features like a variable-speed mixer, precision-controlled fluid circulation, and patented flow stabilization, the Slurry Cart ensures consistent performance and reliable results under HPHT conditions.

Operational Simplicity
This slurry circulation system, in both the 6110 and 6150 configurations, features a 2‑gallon reservoir where test fluids are mixed and maintained in a homogeneous state using a variable‑speed mechanical agitator. For the 6110, a specially designed peristaltic pump circulates fluid through the system after filling, and a 2‑liter accumulator is used to heat and pressurize the circulating test fluid. In contrast, the 6150 utilizes a reciprocating accumulator for circulation, which enables precise flow rate control supported by a patented flow control valve that maintains constant flow during reciprocation. Additionally, the 6150 incorporates a Coriolis flow meter to continuously monitor flow rate and fluid density, providing accurate measurement and system feedback.

Once the system is filled with the test slurry, a closed-loop configuration is established for heating and pressurization. Temperature is regulated using a thermocouple mounted on the accumulator in conjunction with a PID temperature controller, while pressure is applied through a controlled nitrogen source. This setup is designed for seamless integration with the 6100 Formation Response Tester, enabling circulation of drilling fluids through a core sample under controlled thermal and pressure conditions
  • Specifications +


    Features
    • Variable Speed Mixer Keeps Fluids Homogenous
    • PID Controller Provides Precise Temperature Control
    • FRT Software Integrates Data Acquisition with Core Analysis
    • 2 Liter (6110) and 1-1/2 Liter (6150) Piston Accumulator Heats and Pressurizes Circulating Fluid

    Compliance:
    CE
    Environmental: Indoor Use
    Altitude: 6561.6 ft / 2000m
    Temperature: 5 - 40°C / 41 - 104°F
    Relative Humidity: 0% to 95% non-condensing
    Pollution Degree: Category 2

    Operating Conditions
    Reservoir Size: 2 G / 7.6 L
    Maximum Pumping Pressure: 
    6110 -> 1,000 PSI / 6.9 MPa
    6150 -> 5,500 PSI / 37,890 kPa

    System Temperature: 
    6110 -> 20 - 93°C / 68 - 200°F 
    6150 -> 1 - 177°C / 34 - 350°F 

    Maximum Fluid Density: 16.6 lb/gal
    Density Accuracy: 
    0.004 gal/ml (6150)
    Flow Rate Accuracy:
    0.2% (6150)
    Flow Rate: 6110 -> 2.0 LPM / 6150 -> 0.1 - 2.0 LPM
    Air Supply: 
    100 psig / 6.9 BAR, ¼” Swagelok (6150)

    Nitrogen Supply: 
    6110 -> 1,000 psig / 68.9 BAR, ¼” Swagelok
    6150 -> 2,500 psig /103.4 BAR, ¼” Swagelok 

    Water Supply:
    6110 -> Municipal Supply, ¼” Swagelok
    6150 -> 100 psig, / 6.9 BAR, ¼” Swagelok

    Utilities
    Power

    Input Voltage: 
    6110 -> 200 – 240 VAC
    6150 ->208 – 240 VAC

    Input Current: 6110 -> 15A / 6150 -> 20A
    Frequency: 50 / 60 HZ, 1Phase

    Typical Weights and Dimensions
    (w x h x d)
    6110 -> 36 x 72 x 25 in. / 91 x 183 x 64 cm.
    6150 -> 42.5. x 90 x 31.2 in. / 
    108 x 228.6 x 79.2 cm.

    Weight: 985 lb /447 kg



    Manufacturer’s specifications subject to change without notice.




  • Brochures +

  • Standards +


    CE—European Safety Standards